Arterial cannula
Estimated arterial line pressure ≈
Cannula ΔP + arterial tubing pressure loss + patient MAP
Tubing-inlet pressure at patient level, after the oxygenator/filter. Add other downstream losses and correct for height differences.
Compare cannula pressure drop at your target flow using manufacturer curves.
Loading manufacturer pressure-drop references…
Pressure drop data could not be loaded. Please try again.
Scroll the list to browse cannulas. Selected curves appear just below.
Estimated arterial line pressure ≈
Cannula ΔP + arterial tubing pressure loss + patient MAP
Tubing-inlet pressure at patient level, after the oxygenator/filter. Add other downstream losses and correct for height differences.
Resistance to overcome
Cannula |ΔP| + venous tubing pressure loss
Balance against drainage head
CVP · patient-to-reservoir height (Δh) · assisted drainage
Gravity and suction drive drainage against these losses. Cannula ΔP alone is not venous line pressure. CPB reservoir shown; ECMO depends on pump inlet conditions.
Conceptual guide at the same flow; manufacturer curves are reference estimates. CPB fundamentals · Venous pressure interpretation
The dataset includes manufacturer pressure-flow information for selected perfusion cannulas. Availability varies by manufacturer, model, size, and cannula category.
This static indexable summary is generated from the existing cannula pressure-drop dataset source and covers 224 datasets from Getinge / Maquet, LivaNova, Medtronic. Categories represented include Adult arterial, Adult venous, Aortic root / cardioplegia, Cardioplegia, arterial, femoral arterial, femoral bi-caval venous, femoral venous, jugular venous, venous. Representative sizes span approximately 4–51 Fr where French size is listed.
Dataset availability text lists manufacturer, category, model family, and size coverage only; raw pressure-flow point tables remain in the interactive lookup and source dataset rather than displayed as a static data dump.
Pressure-drop estimates are calculated from available manufacturer pressure-flow curves or tables when available, including pressure-flow charts, pressure-drop tables, and product documents for CPB and ECMO cannulae.
When a manufacturer provides only a graph, curve points may be manually digitized and stored as source points. The calculator uses linear interpolation between adjacent source points to estimate ΔP at the selected target flow.
Avalon Elite values were digitized from the manufacturer’s H2O-at-ambient-temperature flow-characteristics graph. Infusion and drainage lumens are shown separately, and pressure values retain their signed source direction; near the zero-pressure axis, digitized drainage points may slightly cross 0 mmHg. Values are reference estimates and should not replace the current manufacturer IFU or institutional monitoring.
If the target flow exactly matches a digitized source point, the result can be labeled as a digitized source point. If the target flow falls outside the available chart range, the tool reports out of source range rather than extrapolating beyond the source data.
Compare at Target Flow applies one shared flow to eligible cannulas across manufacturers and model families, including multiple sizes within the same family. This page does not choose a cannula or define a safe maximum pressure by itself.
Pressure drop increases as flow rises. At the same target flow, a higher pressure drop suggests greater resistance through the selected cannula and related connector geometry.
Comparing nearby sizes, such as 19 Fr versus 21 Fr or 21 Fr versus 23 Fr within the same product family, may help perfusion teams understand relative resistance before CPB or ECMO setup.
Getinge / Maquet HLS cannula entries are commonly interpreted in an ECMO context. When comparing HLS cannula sizes, consider the intended ECMO configuration, cannulation site, drainage or return role, expected flow target, and local institutional practice.
Cannula pressure drop represents the pressure gradient across the selected cannula or manufacturer-tested cannula segment at a given flow. During CPB, measured arterial line pressure is not determined by cannula pressure drop alone and may reflect perfusion circuit resistance plus patient arterial pressure or afterload.
For example, a high arterial line pressure during CPB should not be interpreted as a cannula-size problem automatically. Evaluate the entire arterial pathway, including flow target, MAP/afterload, arterial cannula position, tubing clamps or kinks, connector size, arterial filter or oxygenator pressure gradient, blood temperature, hematocrit, and trends over time.
Interpret the result as a pressure-flow reference, not as an automatic cannula recommendation. Use it alongside target flow, line pressure, drainage or return performance, hemolysis risk, surgical approach, and institutional practice.
Cannula pressure-drop data helps estimate the pressure gradient across a selected cannula at a given flow. In clinical CPB or ECMO, measured circuit pressure reflects the cannula plus the rest of the circuit and patient conditions. Interpret pressure-drop estimates together with arterial line pressure, venous drainage pressure, flow adequacy, reservoir level, patient MAP/CVP, cannula position, and institutional practice.
Arterial pressure monitoring
Arterial cannula pressure test after cannulation
Venous pressure and drainage monitoring
VAVD precautions
How to use this page at the bedside
Manufacturer pressure-flow curves are often generated under controlled bench conditions and may not fully reproduce clinical conditions. Actual pressure drop can vary with blood viscosity, hematocrit, temperature, cannula position, insertion depth, connector size, tubing configuration, kinking or partial obstruction, pulsatility, pump type, and test medium such as water versus blood.
This tool estimates cannula pressure drop from source pressure-flow data. It does not calculate total CPB arterial line pressure and does not account for patient MAP/afterload, oxygenator pressure gradient, arterial filter pressure gradient, tubing resistance, connector resistance, temperature, hematocrit, viscosity, cannula malposition, or obstruction.
Digitized values may include small rounding or measurement errors. Dataset coverage is limited to the currently included manufacturer datasets and may not include every manufacturer, model, size, coating, connector option, or updated product revision.
Values outside the available manufacturer chart range should not be extrapolated. Pressure drop values are reference estimates, not guaranteed patient-specific measurements. Use this tool as a reference alongside manufacturer instructions, patient anatomy, cannula position, blood properties, temperature, flow conditions, tubing/connectors, sensor location, measurement technique, surgical approach, institutional inventory, pressure trends, and local protocols.
What is cannula pressure drop?
Cannula pressure drop is the pressure difference across a cannula at a given flow. It reflects cannula flow resistance and depends on cannula size, geometry, flow rate, fluid properties, and measurement conditions.
How is pressure drop estimated on this page?
The estimate is based on available manufacturer pressure-flow data points. When the entered flow falls between two data points, the page uses linear interpolation between adjacent points. It should be interpreted as a reference estimate for that specific dataset.
Can this calculator be used outside the listed flow range?
No. Estimates should not be extrapolated beyond the available manufacturer data range. If the entered flow is below or above the dataset range, the result should be treated as out of range rather than a reliable estimate.
Does this replace manufacturer instructions or clinical judgment?
No. This tool is a reference aid only. Cannula selection and interpretation of pressure should follow manufacturer instructions, institutional practice, patient anatomy, surgical approach, circuit setup, and clinical judgment.
Why can measured circuit pressure differ from the chart value?
Actual measured pressure can differ because of blood viscosity, hematocrit, temperature, tubing and connector resistance, cannula position, kinking, partial obstruction, vessel size, drainage or return conditions, sensor location, and measurement technique.
Is cannula pressure drop the same as CPB arterial line pressure?
No. Cannula pressure drop is only one component of the pressure observed in an arterial circuit. Measured CPB arterial line pressure may also include patient afterload, oxygenator and arterial filter resistance, tubing and connector resistance, blood viscosity, temperature, hematocrit, cannula position, and circuit configuration.
How should venous pressure-drop data be used?
Venous pressure-drop data can provide a reference for resistance at a target drainage flow, but actual venous return also depends on cannula position, preload/CVP, gravity height, VAVD, tubing configuration, and obstruction.
Each selected lookup preserves the available manufacturer, model family, cannula type, size/code label, flow range, source label, source URL, data status, and digitization note from the pressure-drop dataset.
Public manufacturer references open in a new tab when available. Some products may have source labels or internal dataset notes without a public PDF or product page URL.
Where Medtronic DLP Pediatric One-piece connector/order-code variants share an identical manufacturer Fr-level curve, interactive lookup and comparison show one curve per Fr. Connector specifications and all original product codes remain available in the details and search; this does not imply independent connector-specific pressure testing.
In Compare at Target Flow, independent product curves retain their own source range and status.
The practical monitoring notes summarize general perfusion safety principles from CPB practice guidance, VAVD safety literature, troubleshooting education, and the manufacturer/source pressure-flow data represented in this dataset.
Use Quick Reference for general CPB/ECMO bedside values, Unit Converter for flow and pressure conversions, and BSA Calculator when estimating target perfusion flow before comparing cannula pressure drop.